Refrigerated Centrifuge with Heating System Sigma 6-16KHS
6 Using the centrifuge
32 / 93
Version 10/2012, Rev. 1.19 of 21/03/2019
• sb
Translation of the original operating manual, part no. 0702302
Centrifugation with low capacity
• Install the sample vessels in a rotationally symmetrical manner so that
the buckets and their suspensions are loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 9: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)
Double rectangular bucket 13420 and bucket 13621
The double rectangular bucket 13420 and the bucket for microtiter plates
13621 are dynamically balances at the factory. Therefore, the buckets must
be installed in sets of 4 an in the prescribed order (see the numbers on the
buckets and on the rotor).
CAUTION
Mixed use with only two buckets on opposite sides will cause increased
imbalance!