![Sigma 4-15C Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/sigma/4-15c/4-15c_operating-manual_1267904082.webp)
Operating Manual SIGMA 4-15C, page 82 of 95
03/06
The curve 9 is a special case compared with the other curves. The centrifuge
accelerates with max. torque. The acceleration time only depends on the mo-
ment of inertia of the rotor.
Linear curve no.
Slope
0
240 sec./1000 rpm.
1
180 sec./1000 rpm.
2
120 sec./1000 rpm.
3
60 sec./1000 rpm.
4
40 sec./1000 rpm.
5
30 sec./1000 rpm.
6
20 sec./1000 rpm.
7
10 sec./1000 rpm.
8
5 sec./1000 rpm.
9
0,9 sec./1000 rpm.