Operating Manual SIGMA 4-15C, page 48 of 95
03/06
4.2.7.4 Density
Density
If the density of a sample exceeds 1.2 g/cm
3
, the maximum final speed will
reduce (refer to chapter 9.4.2
“
Density
”
). A value between 1.2 and 9.9 g/cm
3
is
possible.
4.2.7.5 Start Delay
With this function a start delay can be set.
4.2.7.6 Automatic Lid Opening after End of Run
If this parameter has been set the lid will automatically open after standstill of the
rotor.