5.1.3
Alteration of Program Parameters during the Centrifuge Run
During the run, the following values can be altered.
•
Speed
•
RCF
•
Run time
•
Switching into the continuous run mode/time mode
•
Acceleration curve
•
Deceleration curve
•
Start delay
Alter the Set values as usual by activating the Alteration mode (refer to chapter
5.1 "Selection, Display and Alteration of Program Parameters").
You cannot change:
•
Rotor
•
Program
•
Radius
•
Density
The following functions can be activated/deactivated:
•
Start delay
•
Automatic lid opening after end of run