4.2.4
Temperature
The centrifuge is cooled with a cooling system. When a run is started, there is a
temperature difference between the sample and the rotor which is minimized
after a certain operating period has passed.
The scale of the upper display area shows the actual temperature in the
chamber. Activate “set“ in the temperature area using the left knob and select the
set value by turning the right knob. This is possible in steps of 1 °C.
Temperatures between – 10 °C and + 40 °C can be pre selected. If the set
temperature is lower than the actual temperature, the set value is indicated by a
flashing bar.
As soon as the lid is opened, the cooling system is disconnected in order to avoid
icing of the rotor chamber.
Precooling of the rotor below 0 °C during standstil l is not recommended.
Differences between actual and displayed temperature and icing of the
compressor would be the result. This would reduce life of the compressor. We
recommend precooling at 5000 rpm.
The obtainable temperatures depend on rotor type, speed and ambient
temperature. (A change of the limit values might cause deviations.)
Содержание 3-16K
Страница 1: ... Refrigerated Centrifuge 3 16K Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 10: ......
Страница 28: ...Operating Manual SIGMA 3 16K page 28 of 69 03 06 ...
Страница 38: ...Operating Manual SIGMA 3 16K page 38 of 69 03 06 ...
Страница 66: ......
Страница 68: ......
Страница 69: ...Operating Manual SIGMA 3 16K page 69 of 69 03 06 ...