Operating Manual SIGMA 3-16K, page 30 of 69
03/06
1.5
Safety Instructions
According to the German trade association regulation BGR500 chapter 2.11
part 3 the owner of the instrument is advised to take care of the following points:
1.
According to BGR500 the owner has to provide operating instructions
based on those of the manufacturer and to inform the employees
accordingly.
2.
For safety reasons these operating instructions must clearly state that the
stamped max. speed of the used rotor and the max. allowable filling
quantity must not be exceeded.
3.
If the density of the material exceeds 1.2 g/cm
3
, the max. speed of the
centrifuge must be reduced (see formula chapter 7.1.2).
4.
Operation of the centrifuge in hazardous locations is not allowed.
5.
During operation the centrifuge must not be moved. Leaning against or
resting on the centrifuge is not allowed.
6.
Do not spin explosive or highly inflammable materials.
7.
Substances which could damage the material of the centrifuge, the rotors
or the buckets anyhow must not be centrifuged or only under consideration
of special safety measures. Infectious, toxic, pathogene or radioactive
substances must be centrifuged in certified rotors only.
8.
Keep a clearance of at least 30 cm around the centrifuge. Dangerous
materials of any kind must not be put down or stored in that area.
9.
Attention!
Defective lid relieving devices could cause the centrifuge lid to fall down
(contact Service). Risk of bruising!
Содержание 3-16K
Страница 1: ... Refrigerated Centrifuge 3 16K Operating Manual ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 10: ......
Страница 28: ...Operating Manual SIGMA 3 16K page 28 of 69 03 06 ...
Страница 38: ...Operating Manual SIGMA 3 16K page 38 of 69 03 06 ...
Страница 66: ......
Страница 68: ......
Страница 69: ...Operating Manual SIGMA 3 16K page 69 of 69 03 06 ...