5.3.4.11 Data storage mode
a.
Selecting method is the same as “selection freq.&probe” .
b.
Instrument storage method has” manual "and" automatic ",both are effective in the
main measurement interface .Saved information is the current measurement data include"
σ
(20),
σ
(TC), units of measurement, temperature TA and measure frequency f , A
0
value, date” and so on. Save total 16000 sets of measurements data.
c.
Manual save: in the main measurement interface, keep the probe contacting with the
sample after measured test conductivity values, press the storage data key "STORE", at
this time, N=XXX character in the middle of the screen will flash said measurement data
of the page has been saved.
d.
Automatic save: the instrument measure a valid data every time, N=XXX character in
the middle of the screen will flash and save the page of the measurement data
automatically.
e.
Every time shut off and power on again, the instrument will default manual save mode
automatically.
f.
N=XXX numerical is the stored set of memory in the saving process, when full of
16000 sets , the screen shows "FULL" prompt, the user must empty memory then can
continue to implement storage function.( Delete memory method see 5.3.4.9 data query e).
g.
The measurement unit for conductivity value of the saved data is the default unit of
the instrument (see 5.3.4.7), which has nothing to do with the unit “
UNIT
DEL
”selected for the
measurement interface.