- 8 -
temperature compensation-temperature sensor” mode and 6.1.2.3 “automatic
compensation-manual entering temperature” mode).
Examples 3:
continuity measurement operation mode
This mode is only worked under “pseudo temperature compensation” mode, used to select the
conductivity faster, and have alarm function of over upper and lower limitation. Calibrate the high
value and the lower value before measurement (refer to 5.3.3.4~5.3.3.6), setting the upper and
lower limitation of alarm value on measurement interface mode, back to the measurement
interface can measure and select the material.
5.3.2.2 Measuring probe selection
a.
Diameter of material area is larger than 14mm for a planar working piece or concave
surface, select 14m diameter (60KHz) probe. Use the “
” shape hand grip to fix the
inner of the probe.
b.
Measuring the convex surface of the work pieces, select diameter of 14mm (60KHZ)
probe, with "
" type hand grip, adjust the vertical degree of probe plane and convex
surface of work pieces, then tighten the hand grip. Remark: Should calibrate the
instrument before adjusting the hand grip.
c.
Diameter of material is greater than 8mm, select diameter of 8mm (500KHz) probe,
hand with"
" shape grip to push to fasten on the probe of the inner core.
d.
Select diameter of 8mm (500KHz) probe when measure sheet materials (refer to 6.3
material thickness influence), lay down "
" shape grip and fasten the probe surface
on the same plane, increase the stability of the measurement.
e.
It takes 2~5 minutes to warm up after changing the new probe every time. For 500Hz
probe, we suggest to warm up 5 minutes.
5.3.2.3 Measurement interface introduction
After welcome time or in the welcome, calibration and setting interface to press key
"MEAS" directly, the screen appears as one of the following measurement interfaces
(select see 5.3.4.2 measurement interface mode).
or
Explanation:
a
.
The text indicates the function of the interface at the top of the screen,
For example, measuring the conductivity value of the plane specimen is
<measurement
>, Measure camber specimens <measurement
>.