Laboratory Centrifuge Sigma 2-6
6 Using the centrifuge
Version 04/2007, Rev. 2.10 of 06/11/2015 • sb
27
Translation of the original operating manual, part no. 07044
Pos: 95 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 62 Bet rieb _Spinco ntr ol Easy+U nivers al06 2 Bet rieb == == === == == === == == === == == === == == === == == == @ 27\ mo d_14 053 204 171 54_ 68.d ocx @ 194 019 @ 1 @ 1
6 Using the centrifuge
Pos: 96 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 62 Bet rieb _Spinco ntr ol Easy+U nivers al06 2-0 010 Erst e Inb etri ebn ahm e @ 2 7\m od_ 140 532 041 787 5_68 .docx @ 1 940 33 @ 2 @ 1
6.1 Initial start-up
DANGER
• Before the initial start-up, please ensure that your centrifuge is properly
set up and installed (see chapter 5 - "Set-up and connection").
Pos: 97 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 98 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 62 Bet rieb _Spinco ntr ol Easy+U nivers al06 2-0 020 Einsch alten @ 2 7\mo d_1 405 320 418 652 _68. docx @ 19 404 7 @ 2 @ 1
6.2
Switching the centrifuge on
• Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
Pos: 99 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 100 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 062 B etrie b_Spinc ont rol Easy +Unive rsal0 62- 002 0-0 010 Öff nen und Schließ en des De ckels - 2 Schl össer @ 2 7\m od_1 405 320 419 421 _68. docx @ 19 406 1 @ 3 @ 1
6.2.1 Opening and closing the lid
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
• Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
• To close, press with both hands slightly on the lid until both electrical
locks are locked.
WARNING
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!
NOTE
A flashing lid key indicates that only one lid lock has locked.
Pos: 101 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1
Pos: 102 /10 0 Sigm a/10 0 BA Z entrif uge n Sigm a (Sta nda rd mod ule)/ 062 B etrie b_Spinc ont rol Easy +Unive rsal0 62- 002 0-0 020 Einsetz en v on Ro tor en u nd Zub ehö r-- -- --- -- --- -- --- --- -- --- -- --- -- @ 27\ mo d_14 053 204 202 06_ 68.d ocx @ 194 075 @ 3 @ 1
6.2.2 Installation of rotors and accessories
Pos: 103 /10 0 Sigm a/11 0 BA Z entrif uge n Sigm a (Pr ojekt e)/2 -6_ 2-6E/ 062 B etrie b06 2-0 020 -00 20- 001 0 Einset zen des R otor s 2- 6_2 -6E @ 29\ mod _14 0532 122 856 1_6 8.doc x @ 1 964 27 @ 4 @ 1
6.2.2.1 Installation of the rotor
• Open the centrifuge lid by pressing the lid key.
• Unscrew the rotor tie-down screw from the motor shaft (counter-
clockwise).
• Lower the rotor with its central bore straight down onto the motor shaft.
• Tighten the rotor tie-down screw clockwise with the supplied rotor
wrench with 5 Nm. In doing so, hold the rotor at its outer rim.