Laboratory Centrifuge Sigma 2-6
4 Storage and transport
Version 04/2007, Rev. 2.10 of 06/11/2015 • sb
25
Translation of the original operating manual, part no. 07044
Pos: 80 / 100 Sigma /110 BA Ze ntrifu gen Sigma (Proj ekte )/2- 6_2 -6E/0 40 L age run g u nd Tra nspo rt04 0- 0040 Ver pack ung 2-6 _2- 6E @ 2 9\m od_ 1405 321 193 141 _68. docx @ 19 635 7 @ 2 @ 1
s
4.4 Packaging
The centrifuge is packaged in a threepart slip-lid box.
• Take off the lid.
• Remove the box containing the accessories and the packaging material.
• Remove the ring of the slip-lid box.
• Lift the centrifuge upwards with a lifting device or with a sufficient
number of people to lift it safely. When lifting the centrifuge, always
reach under the centrifuge from the side.
CAUTION
The centrifuge weighs approx. 21 kg!
• Retain the packaging for any possible future transport of the centrifuge.
Pos: 81 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 82 / 100 Sigma /110 BA Ze ntrifu gen Sigma (Proj ekte )/2- 6_2 -6E/0 40 L age run g u nd Tra nspo rt04 0- 0050 T ran spo rtsiche ru ng 2 -6_ 2-6E - Übe rsch rift @ 29\ mod _14 053 211 939 00_6 8.d ocx @ 196 371 @ 2 @ 1
4.5 Transport safety device
Pos: 83 / 100 Sigma /110 BA Ze ntrifu gen Sigma (Proj ekte )/2- 6_2 -6E/0 40 L age run g u nd Tra nspo rt/0 40- 005 1 Tra nspo rtsich eru ng 2 -6_ 2-6E - Text @ 36\ mo d_14 346 050 594 89_ 68.d ocx @ 263 827 @ @ 1
The transport safety device consists of a foamed plastic piece in the rotor
chamber.
CAUTION
The transport safety device must be removed prior to start-up!
Removal
• Open the lid by pressing the lid key. If the centrifuge is not connected to
the power supply, use the emergency release of the lid (see chapter
7.1.1 - "Emergency lid release").
• Remove the foamed plastic piece from the rotor chamber, by lifting it
carefully on one side.
• Retain the transport safety device for the possibility of the return of the
centrifuge.