3.6.5 Safety instructions for centrifugation
For safe operation, observe the following before starting the centrifuge:
WARNING
• Ensure that the centrifuge was set up properly (see chapter 5 - "Set-up
and connection").
• Maintain a safety distance of at least 30 cm (12 inches) around the
centrifuge.
• Do not store any dangerous goods in the centrifuge area.
• Do not stay in the safety area longer than what is absolutely necessary
for the operation of the centrifuge.
• Only use the centrifuge with rotors and accessories that have been
approved by the manufacturer. We explicitly warn against the use of
equipment of poor quality. Breaking glass or bursting vessels can
cause dangerous imbalances at high speeds
• Ensure that rotor and buckets are correctly fitted (see chapter 6.2.2.1 -
"Installation of the rotor").
• Observe the instructions on the installation of accessories (see chapter
6.2.2.2 - "Installation of accessories").
WARNING
• The rotor must be loaded axial symmetrically at equal weights.
• If liquids with a density > 1.2 g/cm
3
are used, reduce the speed (see
chapter 2.2.2.2 - "Density").
• Do not use the centrifuge if the rotor is loaded asymmetrically.
• Do not use the centrifuge with tubes that are excessively long.
Pos: 52 / 010 Unive rsalm odul e/Le erzeil e @ 0\mo d_1 202 116 244 500 _0.d ocx @ 114 @ @ 1
Pos: 53 / 100 Sigma /100 BA Ze ntrifu gen Sigma (Stan da rdm odule )/0 30 Sich erh eit03 0-0 060 -00 60 B estän digkeit von K unstst offe n @ 2 6\m od_ 140 531 903 1161 _68 .docx @ 1 921 29 @ 3 @ 1
3.6.6 Resistance of plastics
Chemical influences have a strong effect on the polymeric chains of
plastics, and, therefore, on their physical properties. Plastic parts can be
damaged if solvents, acids, or alkaline solutions are used.
NOTE
• Refer to the resistance data (see chapter 11.4 - "Resistance data")!
Pos: 54 / 010 Unive rsalm odul e/Seite nwechs el @ 0 \mo d_1 202 116 244 312_ 0.d ocx @ 105 @ @ 1