Operating Manual SIGMA 2-5, page 48 of 55
03/06
7.2
Error Correction
Most of the errors can be reset by power off/on. In the event of a short power
failure during a run, this run is interrupted and can be continued by pressing the
Start-key.
No indication on the display:
Actions:
−
No power in the socket?
−
Power cord plugged in and line
voltage present?
−
Input fuse ok?
−
Power switch on?
−
Lid closed?
•
Check fuse in mains supply.
•
Plug in power cord correctly.
•
Replace input fuse (see nameplate for
rating).
•
Switch on power.
•
Close lid (please refer to chapter 7.2.3 “Lid
cannot be opened/closed.“).
7.2.1
Centrifuge cannot be Started
−
Start-key LED not illuminated:
•
Power off/on. If error occurs again, call
service.
−
Lid-key LED flashing:
•
Open and close the lid again. If error
occurs
again
although
both
locks
engaged, call service.
7.2.2
Centrifuge Decelerates during Operation
−
Centrifuge displays an error 1
to 11 after power on.
•
Power off/on. If error occurs again, call
service (please refer to chapter 7.3.1
"Error codes").
7.2.3
Lid cannot be Opened/Closed
−
When first trying to open the lid the locks are not released. The Lid-key LED is
flashing. Open and close lid again.
−
Check/clean the lid seal. Apply talcum powder to the seal to avoid renewed
sticking.
Содержание 2-5
Страница 1: ......
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 10: ......
Страница 18: ...Operating Manual SIGMA 2 5 page 18 of 55 03 06...
Страница 27: ...Operating Manual SIGMA 2 5 page 27 of 55 03 06...
Страница 52: ......
Страница 54: ......
Страница 55: ...Operating Manual SIGMA 2 5 page 55 of 55 03 06...