Refrigerated Centrifuge Sigma 2-16KL
5 Set-up and connection
28 / 82
Version 08/2012, Rev. 1.24 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0700702
Pos : 92 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 050 Aufst ellung und Ans chl uss/ 050 Auf st ellung und Ansc hluss=========================== ======= @ 27\mod_1405319173071_68. docx @ 192381 @ 1 @ 1
→
5
Set-up and connection
Pos : 93 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 050 Aufst ellung und Ans chl uss/ 050- 0010 Aufs tell ort (auß er 8KS) @ 27\m od_1405319174276_68. docx @ 192395 @ 2 @ 1
→
5.1
Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted
to the ambient air.
• Ensure sufficient ventilation.
• Keep a safety distance of at least 30 cm around the centrifuge so that
the vents in the centrifuge remain fully effective.
• Do not subject the centrifuge to thermal stress, e.g. by positioning it
near heat generators.
• Avoid direct sunlight (UV radiation).
• The table must be stable and have a solid, even surface.
• Attention: During transport from cold to warmer places, condensational
water will collect inside the centrifuge. It is important to allow sufficient
time for drying (min. 24 h) before the centrifuge can be used again.
Pos : 94 / 010 U nivers almodule/ Leerz eile @ 0\m od_1202116244500_0. doc x @ 114 @ @ 1
Pos : 95 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 050 Aufst ellung und Ans chl uss/ 050- 0020 Energiev ers orgung------------------------------------------------------ @ 27\mod_1405319175630_68. docx @ 192409 @ 2 @ 1
→
5.2
Power supply
Pos : 96 / 100 Sigma/ 100 BA Zentrif ugen Sig ma (Standar dmodule)/ 050 Aufst ellung und Ans chl uss/ 050- 0020-0011 Ansc hl uss art (Kaltger ät est eck er+Sc hutzsc halter) @ 32 \mod_1417600691402_68.docx @ 222498 @ 3 @ 1
5.2.1
Type of connection
DANGER
The operating voltage on the name plate must correspond to the local
supply voltage!
CAUTION
The mains power plug is an isolating device which is why it must be
accessible at all times.
Sigma centrifuges are units of protection class I. The centrifuges of this
model series have a three-wire power cord with an IEC C13 connector.
NOTE
The removable power cord must not be longer than 3 m!
The power cord must not be replaced with a power cord of inadequate
rating!