Refrigerated Centrifuge Sigma 2-16KL
3 Safety
24 / 82
Version 08/2012, Rev. 1.24 of 28/04/2022
• sb
Translation of the original operating manual, part no. 0700702
Pos : 59 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070 Sic herheits einric ht ung en----------------------------------------- @ 27\mod_1405319034809_68. doc x @ 192157 @ 2 @ 1
→
3.7
Safety devices
Pos : 60 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0010 D eck elverri egelung @ 27\m od_1405319036321_68.docx @ 192171 @ 3 @ 1
→
3.7.1
Lid lock device
The centrifuge can only be started when the lid is properly closed. The
electrical lock must be locked. The lid can only be opened when the rotor
has stopped. If the lid is opened by way of the emergency release system
during operation, the centrifuge will immediately switch off and decelerate
brakeless. If the lid is open, the drive is completely separated from the
mains power supply, i.e. the centrifuge cannot be started (see chapter 7.1.1
- "Emergency lid release").
Pos : 61 / 010 U nivers almodule/ Leerz eile 8pt @ 12 \mod_1337063062114_0. docx @ 65687 @ @ 1
Pos : 62 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0020 Stillst ands überw achung @ 27\mod_1405319037803_68. docx @ 192185 @ 3 @ 1
→
3.7.2
Standstill monitoring system
Opening of the centrifuge lid is only possible if the rotor is at a standstill.
This standstill is checked by the microprocessor.
Pos : 63 / 010 U nivers almodule/ Leerz eile 8pt @ 12 \mod_1337063062114_0. docx @ 65687 @ @ 1
Pos : 64 / 100 Sigma/ 100 BA Zentrif ugen Sigma (S tandar dmodule)/ 030 Sic her heit /030-0070-0031 Syst emk ontroll e Spincontr ol L @ 27\m od_1405319040423_68. docx @ 192213 @ 3 @ 1
→
3.7.3
System check
An internal system check monitors the data transfer and sensor signals with
regard to plausibility. The system continuously performs a self-check and
identifies malfunctions. Error messages are displayed as "Error" followed
by a code number (see chapter 7.2 - "Table of error codes").
Pos : 65 / 010 U nivers almodule/ Leerz eile 8pt @ 12\mod_1337063062114_0. docx @ 65687 @ @ 1
Pos : 66 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0041 Sc hutzleit erpr üf ung ohne Potenti alausgleic hssc hraube @ 27\m od_1405319044593_68. docx @ 192255 @ 3 @ 1
3.7.4
Earth conductor check
An earth conductor check can be carried out by authorised and specialised
personnel using a suitable measuring instrument. Please contact the Sigma
service department (see chapter 7.3 - "Service contact").
Pos : 67 / 010 U nivers almodule/ Leerz eile 8pt @ 12 \mod_1337063062114_0. docx @ 65687 @ @ 1
Pos : 68 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0051 U nw uchtüberwac hung Spi ncontr ol L+Easy+U nivers al @ 27\m od_1405319047031_68.docx @ 192283 @ 3 @ 1
→
3.7.5
Imbalance monitoring system
The indication "Imbalance" in the rotor field and, in some cases, also a
sound signal indicate that the centrifuge is in the impermissible imbalance
range.The drive will be switched off in the acceleration phase or during the
run.
Pos : 69 / 010 U nivers almodule/ Leerz eile 8pt @ 12 \mod_1337063062114_0. docx @ 65687 @ @ 1
Pos : 70 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0060 T emperatur überwac hung KL/KS/Z ent 2 @ 27\m od_1405319048438_68.docx @ 192297 @ 3 @ 1
→
3.7.6
Temperature monitoring system
If the temperature inside the rotor chamber rises above +50°C, the drive
system will be switched off automatically. The centrifuge cannot be
restarted until it has cooled.
Pos : 71 / 010 U nivers almodule/ Leerz eile 8pt @ 12 \mod_1337063062114_0. docx @ 65687 @ @ 1
Pos : 72 / 100 Sigma/ 100 BA Zentrif ugen Sigma (Standar dmodule)/ 030 Sic her heit /030-0070-0070 R ot orüberwac hung @ 27\m od_1405319049932_68.docx @ 192311 @ 3 @ 1
→
3.7.7
Rotor monitoring system
When a rotor number and, if applicable, a bucket number are selected, the
computer will automatically check whether the entered speed or the
entered gravitational field are permissible for the selected rotor.