11.4 Table of the service life of rotors and accessories
• The rotors and accessories must be put out of service after 10 years. Any use after this period
may be permissible in individual cases after an inspection performed by the manufacturer.
• If a specification concerning the maximum number of cycles
and
a specification concerning the
service life (i.e. a date) are provided, the specification that occurs first shall apply.
• After 50,000 cycles, rotors must be scrapped for safety reasons.
Rotor / bucket Cycles Service life
("Exp.Date")
Autoclaving Suitable for centrifuge
Remarks
9100
35,000
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS, 6-16KS,
6-16KHS
9366
15,000
4-5KL, 4-16S, 4-16KS,
4-16KHS, 6-16S, 6-16HS,
6-16KS, 6-16KHS
11805
10 years
8KS, 8KBS
11806
10 years
8KS, 8KBS
12082
7 years
1-14, 1-14K
12083
7 years
1-14, 1-14K
12084
7 years
1-14, 1-14K
12085
7 years
1-14, 1-14K
12092
5 years
20x
1-14, 1-14K
12093
5 years
20x
1-14, 1-14K
12094
5 years
20x
1-14, 1-14K
12096
5 years
20x
1-14, 1-14K
12097
5 years
20x
1-14, 1-14K
12134
5 years
20x
1-16, 1-16K
12135
5 years
20x
1-16, 1-16K
12137
5 years
20x
1-16, 1-16K
12500
7 years
6-16S, 6-16HS,
6-16KS, 6-16KHS
12600
7 years
6-16S, 6-16HS,
6-16KS, 6-16KHS
13218
20,000
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS,
6-16KS, 6-16KHS
13221
10,000
4-16S, 4-16KS, 4-16KHS,
6-16S, 6-16HS,
6-16KS, 6-16KHS
13296
35,000
5 years
10x
2-7, 2-16P, 2-16KL,
2-16KHL
13299
5 years
10x
2-7, 2-16P, 2-16KL,
2-16KHL, 3-30KS,
3-30KHS
13635
25,000
6-16S, 6-16HS,
6-16KS, 6-16KS
13650
20,000
6-16S, 6-16HS,
6-16KS, 6-16KHS
13845
20,000
8KS
13850
10,000
10 years
8KS
13860
15,000
10 years
8KBS
91060
10,000
6-16S, 6-16HS
Special software
required