![Sigma 100008 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/sigma/100008/100008_manual_1267878033.webp)
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL
Refrigerated Centrifuge for Robot Use Sigma 4-5KRL IVD
6 Using the centrifuge
Version 11/2017, Rev. 1.7.1 of 08/12/2022
• sb
33 / 113
Translation of the original operating manual,
→
6
Using the centrifuge
→
6.1
Initial start-up
DANGER
• Before the initial start-up, please ensure that your centrifuge is
properly set up and installed (see chapter 5 - "Set-up and
connection").
→
6.2
Switching the centrifuge on
• Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
NOTE
The start key remains inactive and can only be activated via the
connected PC or by entering a code.
Entering the code
• Press and hold the stop key while entering the code.
• Turn the function knob 1 notch to the left and then 3 nothes to the right.
• Release the stop key.
The start key will be illuminated, the centrifuge is ready for operation.
→
6.2.1
Opening and closing the lid
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
• Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
• To close, press with both hands slightly on the lid until the lock is
audibly locked.
WARNING
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!