Laboratory centrifuge Sigma 1-7
6 Using the centrifuge
Version 08/2018, Rev. 1.7 of 28/04/2022
• sb
35 / 68
Translation of the original operating manual, part no. 0703402
Pos : 133 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0070-0030 Dauerbetrieb @ 27\m od_1405320361344_68.docx @ 193907 @ 4 @ 1
6.3.7.3
Continuous run
During the continuous run, the runtime of the centrifuge is unlimited and
must be stopped manually. The centrifuge accelerates during the
continuous run until the set speed is reached.
• To start the continuous run, press the set key until the time unit flashes
on the display.
• Press the down-key (arrow key) until the display switches from "00:10"
to "--:--".
After 99 min 59 sec, any additional runtime will no longer be displayed, but
the centrifugation will continue.
• To stop a continuous run, press the start/stop key. The centrifugation
will end.
• Enter a runtime. The centrifugation will end after this time.
Pos : 134 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 135 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0090 Softs tart- und Soft st op-F unktionen @ 27\mod_1405320365944_68. docx @ 193949 @ 3 @ 1
6.3.8
Softstart and softstop function
The softstart function is used to extend the acceleration time, whereas the
softstop function is used to extend the deceleration time. The current
combination is shown on the display.
• To activate the softstart and softstop functions, press the set key
repeatedly until "Soft-stop/start" flashes in the upper left part of the
display.
• Press the arrow keys until "Soft-stop/start on" is displayed. The softstart
and softstop functions are now activated.
• Press the arrow key until "Soft stop on" is displayed. Only the softstop
function will be active and the centrifuge will start at normal speed.
• Press the arrow key until "Soft off" is displayed. The softstart and
softstop functions are now deactivated.
The data will be accepted immediately and saved after approx. 20 seconds.
You can also change the settings as described above during a
centrifugation run.