Laboratory centrifuge Sigma 1-7
6 Using the centrifuge
Version 08/2018, Rev. 1.7 of 28/04/2022
• sb
33 / 68
Translation of the original operating manual, part no. 0703402
Pos : 117 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0030 St art en ei ner Z entrif ugati on (außer 2- 7) @ 27\m od_1405320349296_68. docx @ 193795 @ 3 @ 1
6.3.3
Starting a centrifugation run
The centrifuge is ready for operation when the mains power switch is on
and the lid is closed.
• Press the start/stop key in order to start a centrifugation run.
Pos : 118 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 119 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0040 U nt erbr ec hen einer Z entrifugation @ 27\mod_1405320350701_68. doc x @ 193809 @ 3 @ 1
6.3.4
Interrupting a centrifugation run
• Press the start/stop key in order to interrupt a centrifugation run. The
centrifugation run will be terminated prematurely.
Pos : 120 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 121 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0050 U nt erbr ec hen eines Brems vorgangs @ 27\m od_1405320352187_68.docx @ 193823 @ 3 @ 1
6.3.5
Interrupting a deceleration process
• Press the start/stop key during a deceleration process in order to
interrupt it and to restart the centrifuge.
Pos : 122 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 123 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0060 Dr ehzahl / Rel ativ e Z entrifug albesc hleunig ung (RZ B) @ 27\m od_1405320353752_68. docx @ 193837 @ 3 @ 1
6.3.6
Speed / Relative centrifugal force (RCF)
The RCF value is determined by the rotor geometry and speed. The RCF
and speed values, therefore, depend on each other. If one of the two
values is entered, the other value will be set automatically.
• To preselect a speed value or a RCF value, press the set key
repeatedly until the corresponding unit flashes on the display.
• Select the desired speed or RCF value via the arrow keys.
• Confirm the selection by pressing the start/stop key.
The data will also be accepted and saved after approx. 20 seconds if no
other key is pressed within this time period.
During operation, you can switch from the speed value to the RCF value
and vice versa via the arrow keys.
Pos : 124 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 125 /100 Sigma/100 BA Z entrif ugen Sigma (St andardm odul e)/061 Betrieb_Spi nc ontr ol Basic /Spincontr ol Basic/ 061- 0030-0060-0010 Ändern des Drehzahl-/RZ B-Wert es w ähr end des Lauf s @ 27\m od_1405320355244_68.docx @ 193851 @ 4 @ 1
6.3.6.1
Changing the speed/RCF value during centrifugation
The preselected speed or RCF value can be changed during centrifugation.
• Press the set key repeatedly until the desired unit flashes on the
display.
• Change the speed or RCF value by pressing the arrow keys. The
parameters will take effect immediately.
Pos : 126 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1