8 SPD3303S/3303D Quick Start
Output check
The output check includes voltage check in condition of all
channels with no load and current check in condition of short
circuit so as to make sure that the instrument correctly responds to
operation of the front panel.
1. Voltage output check
(1) Within no load, turn on the power, and make sure the setting
current values of all channels are not zero;
(2) Check the voltage output of CH1/CH2
Turn on CH1/CH2 and the instrument work in CV mode. Check
whether the voltage value could be changed from 0V to 32V.
2. Current output check
(1) Turn on the power.
(2) Check the current output of CH1/CH2
Use an insulated wire to connect the positive and negative
terminal of CH1/CH2;
Turn off CH1 and CH2;
Revolve the knob to set the voltage value to 32V;
Revolve the knob to set the current value to 0A;
Adjust the current parameters
to check whether the current
value can be changed from 0A to 3.2A.