If you input a sine signal to channel 1, follow these steps:
1. Measure FFT Amplitude
1) Input a sine signal to channel 1, and press the “AUTO” button.
2) Press the “MATH” button to enter the “MATH” menu.
3) Press the “Operation” option button to select “FFT”.
4) Press the “Source” option button to select “CH1”.
5) Press CH1 button to display CH1 menu.
6) Turn the “S/div” knob to adjust sampling rate (at least double bigger than
frequency of input signal).
7) If FFT display on full screen, press CH1 button again to remove channel
waveform display.
8) Press the “CURSOR” button to enter “Cursor” menu.
9) Press the “Cursor Mode” button to select “Manual”.
10) Press the “Type” option button to select “Voltage”.
11) Press the “Source” option button to select “MATH”.
12) Press the “CurA” option button; turn the “Universal” knob to move Cursor A to
the highest point of the FFT waveform.
13) Press the “CurB” option button, turn the “Universal” knob to move Cursor B to
the lowest point of the FFT waveform.
14) The amplitude (
△
T) displays on the top of the left screen.
Picture 2.6-11
Содержание SDS1000CML+
Страница 1: ...User Manual SDS1000CML SDS1000DL Digital Oscilloscope UM0101A E01A SIGLENT TECHNOLOGIES CO LTD...
Страница 2: ......
Страница 10: ......
Страница 113: ...SIGLENT SDS1000CML SDS1000DL User Manual 103...
Страница 114: ...SIGLENT 104 SDS1000CML SDS1000DL User Manual...
Страница 115: ...SIGLENT SDS1000CML SDS1000DL User Manual 105...
Страница 127: ...Quick Start SDS1000CML SDS1000DL Series Digital Oscilloscope QS0101A E01B 2016 SIGLENT TECHNOLOGIES CO LTD...
Страница 128: ......
Страница 134: ......
Страница 136: ...SIGLENT 2 SDS1000CML SDS1000DL Quick Start Appearance and Dimension Figure 1 Front View Figure 2 Side View...