2

1.  Introduction:

Thank  you  very  much  for  selecting  our  multimeter.  We  are  the    leading 
manufacturer  of  Electrical  and  Electronics  measuring    instruments. These 
multimeters are manufactured as per  IS 13875 and DIN 43751.

2.  Safety features and safety precautions

You  have  chosen  a  multimeter  which  provides  you  a  very  high  degree  of 
safety.  This meter is  manufactured and tested in  compliance   with the safety 
standard IEC 61010-1:2001/ DIN EN61010 -1:2001 and IEC61557.

In case of incorrect use or careless handling, the safety of both user   and 
multimeter is not assured.

For proper use and safe handling, it is absolutely necessary to read  and 
understand the operating instructions before using the meter.

For your safety and for protection of the multimeter, this meter is fitted with an 
Automatic terminal Blocking System  (ABS).

It  is  coupled  with  the  function  selector  switch  which  blocks    the  Terminal 
sockets not necessary for measurement.

Please note the following safety precautions:

l

The multimeter must be operated only by persons who understand the 
danger    of  shock  hazards  and  are  aware  of  the  necessary  safety 
precautions. Shock hazards exist wherever voltages of more than 30V 
(TRMS) are present.

l

Do not work alone in shock hazardous environment while carrying out 
measurement. 

l

The maximum permissible voltage between terminal Socket (7)   and 
ground is 1000 V.

l

Take into account that unexpected voltages can occur on device under 
test  (e.g.  defective  instrument).  For  example,  capacitors  may  be 
charged to a dangerously high voltage.

l

Verify that the test leads are in good condition, e.g.no cracked insulation, 
no open circuits in the leads or connectors.

l

This multimeter must not be used for measurements on circuits with 
corona discharge (high voltage).

l

Be  particularly  careful  when  measuring  on  HF  circuits.  Dangerous 
composite voltages may exist there.

l

Measurements under moist environmental conditions are not Permitted.

l

Do  not  overload  the  measuring  ranges  beyond  their  allowable 
capacities.  Limit values are given in specifications Refer Chapter 17.

l

All current measuring ranges,   are protected with fuse. The maximum   
permissible voltage of the measuring circuit (=nominal voltage of the   
fuse) is1000 VAC/DC in “mA”  ranges.

WARRANTY

Dear Customer,
You are now the privileged owner of 

Zeta 30 

Analog-Digital / Multimeter

 

 a 

product that ranks the first of its kind in the world. 

Company

 provides 12 

months  warranty  from  the  original  date  of  Purchase  against  defective 
material and workmanship.
In  the  unlikely  event  of  failure  of  the  instrument  /  accessories  within  the 
warranty period. 

Company

 will repair meter / accessories   free of charge.   

Please hand over the meter / accessories to the dealer / stockist from whom 
you have purchased along with this card and relevant Cash Memo / Invoice.  
This warranty entitles you to bring the meter / accessories at your cost to the 
nearest stockist / dealer and collect it after repairs.

NO TRANSPORTATION CHARGES WILL BE REIMBURSED.

The warranty is not valid in following cases:

1. Warranty card  duly signed and stamped and original Cash Memo / Invoice 

is not sent along with Meter.

2. Complete warranty card is not presented to authorised person at the time 

of repairs.

3. Meter / accessories is not used as per the instructions in the instruction 

manual.

4. Defect caused by misuse, negligence, accidents, tampering and Acts of 

God.

5. Improper repairing by any person not authorised by the company.
6. Modification, Alteration of any sort is made in electrical circuitry.
7. Seal provided inside/outside is broken.
Warranty of Zeta 20 accessories does not cover Fuses, Battery & Mains 
Adapter.
In case of dispute to the validity of the warranty, the decision of   Company 
service center will be final.
If you bought this Meter directly from the company, and if you notice transit 
damage, then you must obtain the insurance surveyors report and forward it 
to 

Company 

.

Thank you.

(To be filled by authorized dealer)

Model No.

:

Serial No. 

:

Date of Purchase 

:

Cash Memo / Invoice No.  :

Dealer’s Signature 

:

Dealer’s Stamp 

:

Scope of Supply

1) Instrument

2) Cable Set

3) Spare Fuse

4) Safety Cover

5) 1.5 V Battery 6 Nos. - Rechargeable - Yes

No

6) User Manual

7) Warranty Card

8) Crocodile Clips

9) Belt

Содержание Zeta 30

Страница 1: ...rumentation Ltd Central Buildings Woodland Close Old Woods Trading Estate Torquey Devan England TQ27BB Website www sifamtinsley com uk Contact Number 44 O 1803 615139 E mail info tinsley co uk Sifam Tinsley Instrumentation 3105 Creekside Village Drive Suite No 801 Kennesaw Georgia 30144 Contact Number 1 404 736 4903 Web www sifamtinsley com P R E C I S I O N I N S T R U M E N T A T I O N ...

Страница 2: ......

Страница 3: ...shbutton for manual range selection 5 Multi function pushbutton 6 Function selector switch 7 Terminal sockets with automatic blocking system 8 Symbol for CONTINUOUSLY ON 9 Display for digits decimal point and polarity 10 Display for manual range selection DATA hold and MIN MAX storage 11 Display for the selected function 12 Display for the unit of measured quantity 13 Over range indication for pos...

Страница 4: ... measuring adapter 9 Current measurement 9 1 AC current measurement with clip on current transformers 10 Resistance measurement 11 Diode test and continuity test 12 Capacitance measurement 13 Frequency measurement 14 Duty cycle measurement 15 Temperature measurement 16 Insulation resistance measurement 18 1 Battery 18 2 Fuses 18 3 Case 19 Servicing 5 3 Backlit optional 16 1 Before measurement 16 2...

Страница 5: ... capacities LimitvaluesaregiveninspecificationsReferChapter17 l All current measuring ranges are protected with fuse The maximum permissible voltage of the measuring circuit nominal voltage of the fuse is1000VAC DCin mA ranges WARRANTY DearCustomer You are now the privileged owner of Zeta 30 Analog Digital Multimeter a product that ranks the first of its kind in the world Company provides 12 month...

Страница 6: ... Place the multimeter on its face loosen the two screws on the rear and removethelowerpartofthecase liftingitfromthebottom Thelowerandthe upper part of the case are fixed together at the top on the front by means of wedges Ÿ Removeallsixbatteriesfromthebatteryholder Ÿ Placesixnewbatteriesintobatteryholderwithcorrectpolarities Ÿ Replace the lower part of the case Start at the top on the front and t...

Страница 7: ...its for about 10 minutes and when neither a pushbutton nor the function selector switch is operated during that time ItremainsON however whenacurrentmeasuringrangeisselectedand ameasuredvalue 30digitsisdisplayed The meter switches automatically to The next higher range at 3099 digits 1digit the next lower range at 240 280 digits 1 digit from the 300mA to the 3mA range at 24 digits 1 digit 27 Respo...

Страница 8: ... 1999 isshown Manual mode on Used range is fixed 1 x MAN 10 Short Long Return to autoranging 2 X Icap Iea 2 3mA 22 57mA 107 7mA 121 7mA 2 3mA Switching sequence at V 3V 30V 300V 1000V 30mV 300mV 3V 1 x MAN 10 Short Hz 300Hz 3kHz 30kHz 100kHz 300Hz F 30nF 300nF 3 F 30 F 30nF V 3V 30V 300V 1000V 3V mA 300 A 3mA 30mA 300mA 300 A mA 3mA 300mA 3mA 30A 300A 30A 30M 30 300 3 k 30 k 300 k 3 M 30M AUTO MAN...

Страница 9: ...measured value on the digital display and emits a sound signal The probes can now be removed from the measuring point and the measured value on the digital display 9 can be read When the measured value falls below the limit specifiedinthetable themeterisreactivatedforanewstorage The analog indication is not influenced by the DATA hold measured value can still be noted read Note that with a held di...

Страница 10: ...ter acknowledges zero setting by a sound signal the LCD shows 00 00 1 digit and the decimal point flashes The displayed voltage at the instant the pushbutton is pressed is used as reference value max thereafter Thezeroadjustmentisclearedwhen E By pressing the yellow multifunction pushbutton 5 for a long time clearanceisacknowledgedbythetwosoundsignal E ByswitchingtheinstrumentOFF measuringrange 20...

Страница 11: ... 0 1MΩ X V1MW VN V 1000V Continuous 1000V 1000V 1000V 1000V 1000V 1000V max 10S Insulation resistance measurement Measurement Function Measuring Range Resolution Intrensic error of digital display rdg digits at reference conditions V1MW 0 1000V 1V 1 10 MWINSU 1000V 0 1000V 1V 1 10 0 100 1 600 MΩ 01 40 16 00 MΩ 014 0 155 0 MΩ 1KW 10KΩ 1KΩ 100KΩ 5 15 MWINSU Un 50V 0 100 3 100 MΩ 02 80 31 00 MΩ 028 0...

Страница 12: ...ker of 2Aand when thenominal voltageofthesystemdoesnotexceed1000VAC DC l Makethemeasuringcircuitconnectionsmechanicallystrongandsecure so that they do not accidentally open The conductor cross sections and connectionpointsshouldbedesignedtoavoidexcessiveheating l On the 300 mA an Intermittent sound signal warns you when the Measuredvalueexceedstheupperrangelimit l The current measuring ranges upto...

Страница 13: ... z kH 3 00 0 kHz 0 k 10 0 Hz 2 0 98 0 200 0 200 0 0 C 200 0 850 0 0 C 100 0 200 0 0 C 200 0 850 0 0 C fminV 10 pF 2 5 V 100 pF 250 k Ω 2 5 V 300 0 nF 1 nF 25 k Ω 2 5 V 3 000 mF 10 nF 25 k Ω 2 5 V 30 00 mF Icap Iea 2 300 mA 22 57 mA 107 7 mA 121 7 mA 10 pF 100 pF 1 nF 10 nF 250 k 250 k 25 k 25 k 2 5 V 2 5 V 2 5 V 2 5 V Resolution Discharge Resistance U0 max 250 k Ω 30 00 nF Meas urement Func tion M...

Страница 14: ...mA 1mV is used to measurethecurrentupto300AACwiththisfunction E SetrotaryknobatpositionV DC Yellow k measurement with clip on transformer mode E Connect Clamp Output probes to and input terminal of this meter E It has two ranges i e 30 00 A and 300 0A Measurement is possible withboth autorangingandmanualranging AC Pressmultifunction ey until a sound beep is heard This will enter 20 5 Continuously ...

Страница 15: ...10 mV 100 mV 1 V 100 nA 1 mA 10 A m 100 A m 10 mA 100 mA 1 A m 100 A m 10 mW 100 mW 1W 10W 100W 1 kW 10 kW 1 mV 11 M W 40 pF 10 M W 40 pF 10 M W 40 pF 10 M W 40 pF 11 M W 40 pF 10 M W 40 pF 10 M W 40 pF 10 M W 40 pF Voltage drop approx 15 mV 150 mV 650 mV 1V 150 mV No load voltage max 3 2 V max 3 2 V max 1 25 V max 1 25 V max 1 25 V max 1 25 V max 1 25 V max 3 2 V Meas urement Function Measuring R...

Страница 16: ...s on the 30 Ω eliminate the resistance of the leads and contact resistance by zero adjustment range you can Note Resistors and semiconductor junction in parallel with the diode falsify the measuredresults Diodetestandcontinuitytestwithbuzzer With the buzzer function selected the meter emits a continuous sound signalontherange0 approx 0 7V ToswitchtheDiodeTestON Briefly press the yellow multi funct...

Страница 17: ...n device under measurement This position by default reads interference voltage If this voltage is 50V insulationresistancemeasurementisdisabled 16 2 Selecting Test Voltage 50V or 100V or 250V or 500V or 1000V If VINSU key is briefly activated currently selected test voltage is displayed Default values is 500 V To select other value press and hold VINSU key until other voltage is displayed This is ...

Страница 18: ...measurementfunction l by pressing the yellow multi function switch 5 longer this is confirmed bythetwosoundsignals l bychangingthefunctionselectorswitch 15 Temperaturemeasurement The meter Pt100 and Pt1000 temperaturesensorsintherangefrom E Setthefunctionselectorswitch 6 to E Connectthesensortothetwounblockedterminals E Briefly press the yellow multifunction pushbutton 5 The multimeter switches to...

Отзывы: