23
2.
If, upon pressing the
START/STOP
key
(with nominal test voltage set to ≤ 1000V
and load resistance ≥ 5MΩ ), the
capacitance at input leads -Rx, +Rx is >
allowable limit capacitance, the instrument
gives a long acous c signal and displays the
screen reported here to the side for 3
seconds. A erwards, the instrument
returns to the ini al screen.
3.
If, upon pressing the
START/STOP
key
(with 1000V < nominal test voltage set ≤
2500V and load resistance ≥ 5MΩ ), the
capacitance at input leads -Rx, +Rx is >
allowable limit capacitance, the instrument
gives a long acous c signal and displays the
screen reported here to the side for 3
seconds. A erwards, the instrument
returns to the ini al screen.
4.
If, upon pressing the
START/STOP
key
(with 2500V < nominal test voltage set ≤
5000V and load resistance ≥ 5MΩ ), the
capacitance at input leads -Rx, +Rx is >
allowable limit capacitance, the instrument
gives a long acous c signal and displays the
screen reported here to the side for 3
seconds. A erwards, the instrument
returns to the ini al screen.
5.
If, upon pressing the
START/STOP
key or if,
while measuring, an insufficient level of
the supply ba ery voltage is detected, so
that the success of the test is prejudiced,
measurement does not start or is stopped.
The instrument gives a long acous c signal
and displays the screen reported here to
the side for 3 seconds. A erwards, the
instrument returns to the ini al screen.