Series 820 Instruction Manual
Appendix A
IM-82-C
A-1
Appendix A Conversion Formulas and Gas Tables
Conversion of Flow Rate to Other T and P Conditions
The flow rate of your transducer is referenced to certain ÒstandardÓ
conditions of temperature and pressure. Unless otherwise specified in
your order, these standard conditions are 21
°
C (70
°
F) and 760 mm of
mercury (1 atmosphere). If you wish to convert to other ÒstandardÓ
conditions or to find the ÒactualÓ conditions in the pipe where your in-
strument is installed, use the following relationship:
P
1
T
2
Q
2
=
Q
1
(1)
P
2
T
1
( )
1
= The standard conditions under which your instrument
was calibrated,
( )
2
= The new standard conditions or the actual temperature
and pressure conditions in the pipe,
Q
1
=
The gas mass flow rate referenced to the calibrated standard
conditions (sccm or slm),
Q
2
=
The gas mass flow rate referenced to the new standard or
actual conditions (sccm or slmÑÒSÓ means Òstandard,Ó
accm or almÑÒAÓ means ÒactualÓ),
P =
Absolute pressure (kg/cm
2
or psia), and
T =
Absolute temperature (
°
K or
°
R) (
°
K =
°
C + 273,
°
R =
°
F + 460).
Example 1: Changing ÒStandardÓ Conditions
If your transducer has a flow rate reading of 10.00 slm and was calibrated
at standard conditions of 70
°
F (21
°
C) and 1 atmosphere (14.7 psia), and if
you wish to convert this reading to standard conditions of 32
°
F (0
°
C) and
1 atmosphere, then you would use Equation (1) as follows:
14.7
460 + 32
Q
2
=
(10.0) = 9.28 slm
14.7
460 + 70
The flow rate referenced to 0
°
C will be approximately 7% lower than
when referenced to standard conditions of 21
°
C.
Example 2: Finding the ÒActualÓ Flow Rate
If the flow rate and calibrated standard conditions are as given in
Example 1 and you wish to find the actual flow rate at 100
°
F and 30
psig, then you would use Equation (1) as follows:
14.7
460 + 100
Q
2
=
(10.00) = 3.47 slm
14.7 + 30
460 + 70
Содержание Top-Trak 820 Series
Страница 6: ...Table of Contents Series 820 Instruction Manual 0 6 IM 82 C ...
Страница 20: ...Chapter 3 Operation Series 820 Instruction Manual 3 4 IM 82 C ...
Страница 44: ...Series 820 Instruction Manual Appendix B Specifications IM 82 C B 3 ...
Страница 45: ...Appendix B Specifications Series 820 Instruction Manual B 4 IM 82 C ...