50
5
Maintenance
When operated properly and with the use of clean and dry gas,
RedyCompact
devices do not
require any routine maintenance. There are no internal moving parts in the measurement part of
the device. However, we recommend that the calibration is checked after 12 months. If your
RedyCompact
is still within tolerance, this time interval can be extended. The timing of the periodic
check is the customer's responsibility.
5.10.
Cleaning to Remove Contamination
Depending on the type of contamination, on-site cleaning of the measuring or control device may
be possible. As a first step, we recommend flushing with N
2
or dry air. If it is contaminated with
liquids (ex. oil), pure methanol alcohol (100%) can be used. After cleaning, rinse the
RedyCompact
with the optional needle valve position fully open with dry air or nitrogen for approximately 15
minutes to dry out all liquids.
Flow Splitter Disassembly
If the basic body has become contaminated, the flow splitter can be removed. The removal should
only be done by trained service personnel. Removal and Assembly can potentially shift the
calibration of a unit. Disassembly is different for the different device types:
Note: The fourth letter of the article code defines the type of the flow splitter. For example:
GCM-B9EA-BB20 contains a flow splitter of the type B.
Type A
•
First release the slotted screw in the center of the flow splitter (Left tread, approx. 5
turns)
•
Unscrew the whole flow splitter with a hex key
Warning
•
The warranty is null and void if the housing was removed.
•
Only use the proper tools.
•
Be careful when handling the device and the individual components.
•
Make sure that the disassembly environment is clean.
•
Do not touch the circuit board or electronic components without first
grounding yourself and the surroundings. Electrostatic discharges can
destroy components.
•
After cleaning, you should have the device checked or, if necessary,
recalibrated by your sales partner at the next opportunity.
Warning
Before you carry out any mechanical work on the unit the gas supply must be
disconnected, any pressure relieved and the all power disconnected/removed