Chapter 5 Troubleshooting and Service
Series 810 Instruction Manual
Controller does not
Low or no gas pressure
respond to set point
Faulty cable or connector
Set point is below 2% of full scale
DIP switches improperly set
Flow does not match set
point
Reads full scale with no
flow or with valve shut
No gas pressure (or low pressure)
Inlet filter screen clogged
Out of adjustment
Gas Leak
Defective Sensor
Out of calibration
Dirty or clogged sensor
Change in composition of gas
Gas leak
PCB defective
LFE dirty
Inlet filter screen clogged
Reference conditions incorrect
Set correct gas pressure
Correct or replace
Increase set point or disable auto
shut off circuit
Re-configure DIP switches
Set correct gas pressure
Clean or replace
Consult factory
Find and correct leaks
Return to factory for replacement
Clean or replace sensor
See K-factor tables
Find and correct leaks
Repair or replace PCB
Consult factory
Clean or replace screen
Check reference conditions on data
label and correct as required.
5-2
IM-81