Head
Base
Cerebrum
kV
130
130
Effective mAs
240
300
Slice collimation 1.0 mm
2.0 mm
Slice width
4.0 mm
6.0 mm
Feed/Rotation
4.0 mm
8.0 mm
Rotation time
1.0 sec.
1.0 sec.
Kernel
H31s
H31s
Increment
4.0 mm
6.0 mm
Image order
ca-cr
ca-cr
CTDI
Vol
65.9 mGy
70.7 mGy
PosteriorFossaOptimization Filter
PFO: To reduce beam-hardening artifacts in head
images, use for both ranges the Posterior Fossa
Optimization (PFO) filter.
Contrast medium IV injection
Start delay
60 sec.
Flow rate
2 ml/sec.
Total amount
50 – 60 ml
85