60
AbdPelRoutine
Indications:
Spiral mode for routine studies in the region
of abdomen and pelvis, e.g. tumors, lymphoma,
abscesses, post-traumatic changes, etc.
A range of 30 cm will be covered in 15 sec.
Abd
kV
110
Effective mAs
*
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
18.0 mm
Rotation time
0.8 sec.
Kernel
B41s
Increment
5.0 mm
Image order
cr-ca
* The mAs should be adjusted to the body weight
of the child.
Children
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1 – 2 ml per kg of body weight
Body weight
kV
mAs
CTDI
Vol
(mGy)
< 15 kg
110
15
1.2
15 – 24 kg
110
30
2.4
25 – 34 kg
110
50
3.9
35 – 44 kg
110
75
5.8
45 – 55 kg
110
105
8.2