159
AngioFast
kV
110
Effective mAs
90
Slice collimation
2.0 mm
Slice width
3.0 mm
Feed/Rotation
21.0 mm
Rotation time
0.6 sec.
Kernel
B31s
Increment
3.0 mm
Image order
cr-ca
CTDI
Vol
7.0 mGy
Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.
* CARE Bolus may be used to optimize the bolus timing.
Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.
Contrast medium IV injection
Start delay
20 sec.*
Flow rate
3.5 ml/sec.
Total amount
120 ml
Abdomen
MIPthin: AngioFast
coronal
Image thickness
3
Distance between images
3
Number of images
35
For the 1
st
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Содержание SOMATOM Emotion 6
Страница 1: ...SOMATOM Emotion 6 Application Guide Routine Protocols Software Version A70 ...
Страница 23: ...23 General ...
Страница 43: ...43 Children ...
Страница 59: ...59 Children ...
Страница 61: ...61 Children Tips Children with a body weight of more than 55 kg should be examined with an adult protocol ...
Страница 63: ...63 Children Tips Children with a body weight of more than 55 kg should be examined with an adult protocol ...
Страница 89: ...Head 89 ...
Страница 102: ...102 Shoulder Overview Shoulder Spiral mode for HiRes bone studies and soft tissues ...
Страница 139: ...139 Abdomen ...
Страница 167: ...167 Pelvis ...
Страница 175: ...175 Spine ...
Страница 185: ...185 Extremities ...