Commissioning
SITRANS WF100
28
Operating Instructions, 02/2012, 7ML19985NB01-01
For example:
If the design flowrate is 6 T/H, then:
Test weight 60% = 6 t/h × 60% × 45 g/t/h
= 162 g
Test weight 80% = 6 t/h × 80% × 45 g/t/h
= 216g
Therefore the test weight should be between 162 and 216 g.
If an available test weight of 200 g is used, then:
Test rate = 200 g ÷ 45 g t/h
= 4.444 t/h
Perform the zero calibration as described in the Calibration section of the integrator manual.
After the completion of the zero calibration, a span calibration, as described in the integrator
instruction manual, can be performed with the test weight(s) applied.
Span
1.
Install the calibration pulley to the bracket on the flowmeter housing.
2.
Hang the test weight over the calibration pulley using the test weight cable.
3.
Perform the span calibration as described in the calibration section of the integrator
instruction manual.
Содержание SITRANS WF100
Страница 1: ...SITRANS Solid Flowmeters SITRANS WF100 Operating Instructions 02 2012 ...
Страница 6: ...Table of contents SITRANS WF100 4 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 8: ...Introduction SITRANS WF100 6 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 22: ...Description 3 1 Material feed examples SITRANS WF100 20 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 26: ...Installing Mounting SITRANS WF100 24 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 28: ...Connecting SITRANS WF100 26 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 32: ...Commissioning SITRANS WF100 30 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 42: ...Technical data SITRANS WF100 40 Operating Instructions 02 2012 7ML19985NB01 01 ...
Страница 44: ...Dimension drawings SITRANS WF100 42 Operating Instructions 02 2012 7ML19985NB01 01 ...