Installing/Mounting
SITRANS WF100
22
Operating Instructions, 02/2012, 7ML19985NB01-01
Load cell handling
WARNING
Negative displacement on the load cell can functionally destroy the load cell. When
handling the load cell, care must be taken not to force the gap open.
Figure 4-1
Support load cell throughout handling in non-vertical positions
Figure 4-2
Improper handling may functionally destroy the load cell
Installation
1.
Set the flowmeter housing in place ensuring that the flowguide and outlet gaskets are
correctly located (gaskets supplied by customer).
2.
The housing must be vertical within 1° in two planes 90° apart.
3.
Bolt the flowguide and outlet flanges to the process piping.