
Troubleshooting/FAQs
7.2 Fluctuating process values
35
7.2
Fluctuating process values
Question
Why do the displayed process values fluctuate when the electrode cable is moved?
Answer
There are several causes to the fluctuating process values:
●
Deposits on electrodes
–
Clean the electrodes.
●
Defect electrode cable
–
Replace the cable
●
Incorrect cable connection
–
Connect the electrode cable (82, 83, 0 and shield) according to the instructions in
chapter Connecting (Page 23)
NOTICE
Vibrating environments
It is recommended to use special low noise cables for sensor sizes DN 2 and 3
installed in vibrating environments.
www.eltra-trade.com [email protected]
+421 552 601 099
Содержание SITRANS MAG 5100W
Страница 1: ...SITRANS FM MAG 5100 W Manual ...
Страница 16: ...Description 3 2 Theory of operation 14 www eltra trade com info eltra trade com 421 552 601 099 ...
Страница 52: ...Technical data 8 8 Dimensions and weight 50 www eltra trade com info eltra trade com 421 552 601 099 ...
Страница 60: ...Appendix A 4 Ordering 58 www eltra trade com info eltra trade com 421 552 601 099 ...
Страница 62: ...Glossary 60 www eltra trade com info eltra trade com 421 552 601 099 ...