Connecting
The following contains a short description of how to connect a remote mounted sensor to a
transmitter type SITRANS F M MAG 5000 / 6000 or MAG 6000 I. For more information, e.g.
about wiring of power supply and outputs, refer to the Operating Instructions for the
respective transmitters.
Before connecting
●
Check that serial numbers on sensor and SENSORPROM® unit are identical.
WARNING
The pertinent regulations must be observed for electrical installation.
Never install the device with the mains voltage switched on!
Danger of electric shock!
The electrodes and magnetic current line may only be connected when the device is not
connected to the power supply.
If the housing is under voltage (power supply), the cover may be unscrewed by qualified
personnel only.
WARNING
Mains supply from building installation Class II
A switch or circuit breaker (max. 15 A) must be installed in close proximity to the equipment
and within easy reach of the operator. It must be marked as the disconnecting device for
the equipment.
Cable specifications
●
Only use cables with at least the same degree of protection as the sensor to install the
sensor.
●
The line length from the cable gland to the terminals must be kept as short as possible.
Line loops in the terminal box must be avoided.
www.eltra-trade.com [email protected]
+421 552 601 099
Содержание SITRANS MAG 5100W
Страница 1: ...SITRANS FM MAG 5100 W Manual ...
Страница 16: ...Description 3 2 Theory of operation 14 www eltra trade com info eltra trade com 421 552 601 099 ...
Страница 52: ...Technical data 8 8 Dimensions and weight 50 www eltra trade com info eltra trade com 421 552 601 099 ...
Страница 60: ...Appendix A 4 Ordering 58 www eltra trade com info eltra trade com 421 552 601 099 ...
Страница 62: ...Glossary 60 www eltra trade com info eltra trade com 421 552 601 099 ...