14
SFIDK.PS.022.Q5.02
Install the pulse cable through the glands,
before pushing
the printed circuit board back into
place in the base part.
For compact version, install the pulse cables and power supply cable through the gland without
remove the printed circuit board.
For FUS380 and FUE380 the pulse output A and B settings are depending form the ordering:
Normal factory setting see following table. The settings for FUS380 can be read out and changed
with a PC, the software SIMATIC PDM (Process device manager) and the IrDa optical interface
(see accessories for FUS380 in FI01 catalogue). The approved FUE380 settings cannot be
changed according the approval requirements, but it can be read out per PDM.
Electrical connection
5
5.1
Pulse output A and B setting
FUS380
FUE380
Output A
Forward / reverse
Preset: Forward
Output B
Preset: Alarm
Preset: Alarm
Output B, function
Reverse pulse, alarm, call-up
Preset: Alarm
Pulse value A & B
0.1 l/p; 0.25 l/p; 0.5 l/p; 1 l/p; 2.5 l/p;
Preset: See scheme to FUE380.
(depenting on DN value)
10 l/p; 25 l/p; 50 l/p; 100 l/p; 250 l/p;
500 l/p; 1 m3/p; 2.5 m3/p;
Settings typically for
5 m3/p; 10 m3/p; 25 m3/p;
SITRANS FUE950 energy
50 m3/p; 100 m3/p;
calculator.
250 m3/p; 500 m3/p;
1000 m3/p
Pulse with
5; 10; 20; 50; 100; 200; 500 ms
Preset: 5 ms
5.1.1 FUE380 preset pulse output A settings dedicated to energy
calculator type SITRANS FUE950
Output A,
terminals 56/57:
Pulse rate can be seen on the transmitter side label and
must correspond
to the pulse setting
of the energy calculator type.
Factory pre-settings SITRANS FUE380. (Pulse width 5 ms).
DN
Pulse setting (liter/pulse )
50
1
65
1
80
2.5
100
2.5
125
2.5
150
10
200
10
250
10
300
50
350
50
400
50
500
100
600
100
700
100
800
100
900
100
1000
100
1200
100