Adjustments / Programs
5 - 11
Siemens AG
Register 5
RXR2-130.061.02
Page 11 of 22
SIREMOBIL Compact
Medical Engineering
Rev. 03
10.03
CS PS 24
System Manual
•
Release fluoroscopy. The X-Iris must close between 4 - 4.5 cm.
•
Stop fluoroscopy.
•
Select "5. CLOSED POSITION" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "No".
•
Select "8. Init Off & Store Positions" on the service PC and click on "Execute".
•
Close the status window by clicking on the OK button.
•
For the query "Do you want to reset the unit," click "Yes".
•
Switch the SIREMOBIL Compact OFF and ON again.
•
Check the X-Iris again in survey format and zoom format. The collimator blades must still
be visible on the edge of the image. If necessary, repeat the adjustment.
•
Stop fluoroscopy.
•
Return to the main menu on the service PC with the ESC key.
•
Select the Data, Backup... menu.
•
Select Parameter in the backup window.
•
Click on "Backup".
•
At "Remarks for Parameter" enter "X-Iris set“.
•
Click on "OK". The new position values for the collimator will be stored on the disk.
•
Return to the main menu with the ESC key.