Siemens AG
RA52-060.061.01.07.02
SIMOMED HM
07.02
TD SD 21
Configuration with Monitor Replacement
15
Page 15 of 24
Medical Solutions
Jumper Configuration (Hardware)
0
NOTE
If one of the jumper positions described below is changed, the
monitor must be disconnected from line power!
Jumpers for Monitor Size
Video Terminal Resistor / Cable Compensator
•
Remove the rear wall of the housing
•
Remove the top cover panel.
Multistar KK with
display 2000 60Hz
Norm 7 (2045
lines/120Hz/1:1)
Norm 8 (1023 lines/120Hz/4:3)
AXIOM Artis Bypass (!)
Norm 3 (1125
lines/60Hz/1:1)
n.a.
System
Norm Used
Blocked Norm (black background)
Behind the front plane, on the digital
controller, there are two jumpers:
•
X11 => configuration of the monitor
size. Incorrect configuration leads to
the error: EEPROM (in this case, re-
position the jumper).
- 1 installed => 44cm (17”)
- not installed => 54cm (21”)
•
X8 => Sensitivity of the ambient light
sensor (factory setting, 2-3 middle).
Fig. 2:
Monitor size
Remove the 2 screws on the rear wall, take
the housing off towards the back.
Fig. 3: