Configuration with Monitor Replacement
SIMOMED HM
RA52-060.061.01.07.02
Siemens AG
07.02
TD SD 21
Page 14 of 24
Medical Solutions
System
Norm Used
Blocked Norm (black background)
Hicor / ACOM 50Hz
Norm 12 (1249
lines/100Hz/4:3)
Norm 11(2497 lines/100Hz/1:1)
HICOR / ACOM 60Hz
Norm 8 (1023
lines/120Hz/4:3)
Norm 7 (2045Hz/120Hz/1:1)
HICOR 50Hz Bypass (!)
Norm 5 (1249
lines/50Hz/1:1)
n.a.
HICOR 60 Hz Bypass (!) Norm 1 (1023
lines/60Hz/1:1)
n.a.
Multistory TOP
Neurotic TOP
AngiostarTOP
Norm 10 (2251
lines/120Hz/3:4)
n.a.
MST TOP / NST TOP /
AST TOP
=> in Bypass (!)
Norm 3 (1125
lines/60Hz/1:1)
n.a.
Fluorospot TOP 1K
AXIOM Artis ....
Norm 2 (1066
lines/72Hz/5:4)
n.a.
Fluorospot TOP 2K
Norm 9 (1066
lines/76Hz/5:4)
n.a.
Fluorospot TOP Bootup
(!)
Norm 1 (1023
lines/60Hz/1:1)
n.a.
Fluorospot Compact
Norm 2 (1066
lines/72Hz/5:4)
n.a.
Leonardo
Norm 2 (1066
lines/72Hz/5:4)
n.a.
Fluorospot H 50Hz
Norm 5 (1249
lines/50Hz/1:1)
n.a.
Fluorospot H 60Hz
Norm 1 (1023
lines/60Hz/1:1)
n.a.
Multistar KK without
display 2000 50Hz
Norm 5 (1249
lines/50Hz/1:1)
n.a.
Multistar KK without
display 2000 60Hz
Norm 1 (1023
lines/60Hz/1:1)
n.a.
Multistar KK with
display 2000 50Hz
Norm 1 (12497
lines/100Hz/1:1)
Norm 12 (1249 lines/100Hz/4:3)