System diagnostics
9.2 XML/PLC error messages
SIMATIC RF600
478
System Manual, 06/2019, J31069-D0171-U001-A22-7618
"ER"
LED
XML/
LED
(hex)
PLC block
(hex)
Error description
--
0x65
0xE6FE05 An error has occurred that makes a Reset_Reader ("WRITE-CONFIG" with "Config = 3")
necessary.
Possible causes / action to be taken:
•
The "WRITE-CONFIG" command is incorrect.
•
After eliminating the error, execute an "INIT".
•
Check the "IID_HW_CONNECT" parameter.
--
0x66
0xE6FE06 The reset timer has expired.
2 Hz
0xE1 0xE6FE81 A parameter is missing.
2 Hz
0xE2 0xE6FE82 The parameter has an invalid format.
2 Hz
0xE3 0xE6FE83 The parameter type is invalid.
2 Hz
0xE4 0xE6FE84 Unknown parameter.
2 Hz
0xE5 0xE6FE85 The command or the frame has an invalid format.
2 Hz
0xE6 0xE6FE86 The inventory command failed.
2 Hz
0xE7 0xE6FE87 Read access to the transponder failed.
2 Hz
0xE8 0xE6FE88 Write access to the transponder failed.
2 Hz
0xE9 0xE6FE89 Writing the EPC-ID on the transponder failed.
2 Hz
0xEA 0xE6FE8A Enabling write protection on the transponder failed.
2 Hz
0xEB 0xE6FE8B The "Kill" command failed.
2 Hz
0x71
0xE7FE01 In this status, only the "Reset_Reader" command ("WRITE-CONFIG") is permitted.
--
0x72
0xE7FE02 The "CMD" command code is not permitted.
--
0x73
0xE7FE03 The "LEN_DATA" parameter of the command is too long and does not match the global
data reserved within the send data buffer (TXBUF).
--
0x74
0xE7FE04 The receive data buffer (RXBUF) or the send data buffer (TXBUF) is too small, the buffer
created at TXBUF/RXBUF does not have the correct data types or the parameter
"LEN_DATA" as a negative value.
Possible cause / action to be taken:
•
Check whether the buffers TXBUF/RXBUF are at least as large as specified in
LEN_DATA.
•
With S7-1200/1500:
–
In the Ident profile, only an "Array of Byte" may be created for TXBUF and RXBUF.
–
In the "Reader_Status" block, only an "Array of Byte" or the corresponding data
types ("IID_TAG_STATUS_XX_XXX" or "IID_READER_STATUS_XX_XXX") may be
created
--
0x75
0xE7FE05 Error message that informs you that only an "INIT" command is permitted as the next com-
mand. All other commands are rejected.
--
0x76
0xE7FE06 Wrong index
Permitted index is in the ranges "101 ... 108" and "-20401 ... -20418".
--
0x77
0xE7FE07 The reader or communications module does not respond to "INIT" ("INIT_ACTIVE" is ex-
pected in the cyclic status message).
The next steps:
•
Check the address parameter "LADDR".
--
0x78
0xE7FE08 Timeout during "INIT" (60 seconds according to "TC3WG9")