Set CEM CERT
Operating Instructions, 06/2021, A5E34491309-12
107
Technical specifications
11
System design
Color
RAL 7035
Weight
Approx. 160 kg (353 lbs.)
Sheet-steel cabinet/frame
Dimensions (with base)
1)
Indoor installation
2100 x 800 x 800 mm; 83" x 31 1/2" x 31 1/2" (H x W x D)
GRP cabinet
Dimensions (with base)
1)
Outdoor installation
2060 x 900 x 800 mm; 81" x 35 1/2" x 31 1/2" (H x W x D)
Explosion protection classification
Installation outside the Ex zone
Degree of protection
IP 54 (avoid direct sunlight when installed outdoors)
Operating position
Vertical
Sample gas cooler
2-stage
Calibration
•
ULTRAMAT 23
•
ULTRAMAT 6, OXYMAT 6
Semi-automatic; AUTOCAL at 24-hour intervals
Fully automatic
Analyzers
ULTRAMAT 23, ULTRAMAT 6, OXYMAT 6 and SIPROCESS UV600
For further technical details of the analyzers, refer to the
corresponding operating instructions. This information is included in
section 4 of the overall documentation with your delivery.
1)
A space of 500 mm (20") must be provided on the right or left side for the cable inlet and the connection of the
heated sample gas line.
Climatic conditions
Permissible ambient temperature
•
During operation
•
With heating in sheet-steel cabinet
•
With heating in GRP cabinet
•
With air conditioning
+5
°C ... +40 °C, standard; conforms to EN 15267
-3
-5
°C ... +45 °C
-15
°C ... +40 °C
Max. +52
°C
Permissible ambient humidity
75% RH (relative humidity) annual average, non-condensing
The conditions of EN 15267-3 must be observed.
Содержание Set CEM CERT
Страница 1: ......
Страница 2: ......
Страница 8: ...Table of contents Set CEM CERT 6 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 24: ...Description 3 3 Principle of operation Set CEM CERT 22 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 26: ...Application planning 4 2 Checking the delivery Set CEM CERT 24 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 36: ...Connection 6 2 Electrical installation Set CEM CERT 34 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 44: ...Commissioning 7 2 Commissioning Set CEM CERT 42 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 62: ...Operation 8 4 Menu based operation Set CEM CERT 60 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 88: ...Operation 8 4 Menu based operation Set CEM CERT 86 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 89: ...Operation 8 4 Menu based operation Set CEM CERT Operating Instructions 06 2021 A5E34491309 12 87 ...
Страница 91: ...Operation 8 4 Menu based operation Set CEM CERT Operating Instructions 06 2021 A5E34491309 12 89 ...
Страница 106: ...Service and maintenance 9 6 Spare parts accessories Set CEM CERT 104 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 112: ...Technical specifications Set CEM CERT 110 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 116: ...Dimension drawings 12 1 Dimension drawings Set CEM CERT 114 Operating Instructions 06 2021 A5E34491309 12 ...
Страница 117: ...Set CEM CERT Operating Instructions 06 2021 A5E34491309 12 115 Wiring diagrams 13 13 1 Gas flow diagram ...
Страница 122: ...Appendix A 2 Returns Set CEM CERT 120 Operating Instructions 06 2021 A5E34491309 12 ...