
POWERMOBIL
SPR2-240.841.01
Page 2 of 6
Siemens AG
Rev. 01
05.03
CS PS 24
Medical Solutions
4 - 2
Checks and Adjustments
•
Keep the C-arm in this position.
•
Turn the bell (1/Fig. 5) until the rollers (2/Fig. 5) are aligned with the marking on the bell
(3/Fig. 5). Fig. 6 shows the bell (1/Fig. 6), the marking (3/Fig. 6) and the shaft with the two
rollers (2/Fig. 6) once again in a removed state.
•
In this position of the bell and rollers, push the pinion gear back into the original position.
Make sure that the teeth of the gear engage.
•
Retighten the screw (1/Fig. 3).
Fig. 5
Fig. 6