10.
Take either transducer and apply a continuous lengthwise 1/8-inch bead of coupling
compound across the center of the transducer emitting surface.
①
Coupling Compound
④
Transducer
②
Emitting Surface
⑤
Angled Edge
③
Back End
⑥
F-Connector
Figure 3-6
Transducer
11. Slide transducer into a Mounting Frames back end first aligning the angled edge of the
transducer with the angled edge of the Mounting Frame. Keep transducer from making
contact with the pipe until it butts up against the Mounting Frame stop.
12.
Tighten the transducer Clamping Screw to hold the transducer in place.
Repeat
installation procedure for the other transducer.