CR-IR347
Service Manual
PM - 14
009-058-03
08.30.2002 FM3476
PM - 14
■
Check for Jitter and HR
(1) Place steel rules and a washer on IPs of all sizes found at the institution, and expose
them. (Both ST and HR types. For the layout, see the illustration below.)
The IP exposure conditions are as follows.
❍
ST
• Dose:
1 mR
• X-ray tube target :
Tungsten
• Distance:
1.8 m
• Voltage:
80 kV
• Current:
50 mA
• Time:
0.013 sec
❍
HR
• Dose:
20 mR
• X-ray tube target/filter: Molybdenum/molybdenum
• Distance:
0.55 m
• Voltage:
28 kV
• Current:
100 mA
• Time:
0.053 sec
The layout of the steel rules and washer is shown below.
FR7H6950EPS
Cassette
Steel rules (150mm)
Steel rules (150mm)
Cassette
Washer (ø8–15)
Washer (ø8–15)
14" x 17"
24cm x 30cm mammo
Содержание CR-IR 347
Страница 2: ...CR IR347 CR IR347P Service Manual Preventive Maintenance PM ...
Страница 22: ...CR IR347 Service Manual PM 15 3 009 058 03 08 30 2002 FM3476 PM 15 3 BLANK PAGE ...
Страница 23: ...CR IR347 Service Manual PM 15 4 009 058 03 08 30 2002 FM3476 PM 15 4 BLANK PAGE ...
Страница 29: ...CR IR347 Service Manual PM 21 009 058 00 10 20 2000 FM2732 2 PM 21 BLANK PAGE ...
Страница 59: ...CR IR347 Service Manual PM 51 009 058 00 10 20 2000 FM2732 2 PM 51 BLANK PAGE ...
Страница 73: ...CR IR347 Service Manual PM 65 009 058 00 10 20 2000 FM2732 2 PM 65 BLANK PAGE ...
Страница 83: ...CR IR347 Service Manual PM 75 009 058 00 10 20 2000 FM2732 2 PM 75 BLANK PAGE ...
Страница 89: ...CR IR347 Service Manual PM 81 009 058 00 10 20 2000 FM2732 2 PM 81 BLANK PAGE ...
Страница 110: ...CR IR347 Service Manual PM 102 009 058 00 10 20 2000 FM2732 2 PM 102 BLANK PAGE ...
Страница 119: ...CR IR347 Service Manual PM 111 009 058 00 10 20 2000 FM2732 2 PM 111 BLANK PAGE ...