Requirements
1 - 9
Siemens AG
SPR8-125.034.02
Page 9 of 10
POLYMOBIL Plus
Medical Solutions
Rev. 02
03.05
CS PS 24
Replacing damaged or missing screws
1
•
As specified in the installation drawings, damaged or missing screws may only be
replaced by steel screws conforming to DIN 267 and having the specified tensile
strength.
Cleaning
1
•
The unit must always be switched off or disconnected before cleaning. Use only water to
clean the unit or a lukewarm solution consisting of water and a household cleaning
agent.
•
Do not use abrasive cleaning agents or organic solvents such as benzine, alcohol or spot
remover.
•
Do not spray water on the unit.
•
For additional information, refer to the Instructions for use, chapter "Cleaning/
Disinfecting".
For the assessment the first measured value and the values docu-
mented during previous maintenance or safety checks must be
compared with the measured values.
An abrupt rise of the measured values, even if the limit of 2.5 mA
is not exceeded, indicates defects in the primary circuit of the
mains voltage supply (insulation damage, moisture damage,
defective interference suppressor, etc.).
Replacement of screws.
All Allen screws must have a tensile strength rating of 8.8.
NOTE