32
8015383/ZUQ7/V1-3/2020-06 | SICK
O P E R A T I N G I N S T R U C T I O N S | TRANSIC151LP
Subject to change without notice
4
OPERATION
4
Operation
4.1
Safety instructions for operation
4.2
Interfaces
There are 2 control interfaces
●
Keypad (on the front TRANSIC151LP panel)
●
Maintenance interface
4.2.1
Control via keypad
A display and four pushbuttons are located on the front enclosure panel. The measured oxy-
gen value is shown on the display. LEDs signal the operating mode of the measuring device.
A green LED indicates normal operation.
Characteristics
The main purpose of the integrated interface (keypad/display) is field calibration.
The following values can be set to attain higher measurement precision:
●
Process pressure
●
Humidity
●
Carbon dioxide content
NOTE:
Risk of damage to the TSA151 through dust or humidity
▸
Only open the TSA151 in a dry environment free from dust.
NOTE:
Read the instructions through carefully before making any settings or
parameter changes. SICK accepts no responsibility for parameter or setting
changes nor adjustments made by the user. Contact SICK Customer Service
when you require technical support.
The password can be found in the Annex,
.
DANGER: Hazard through incorrect parameter settings
Incorrect settings for parameters can have severe consequences. This is why the
password may only be available for authorized technicians.
▸
Note the password in the Manual and keep it safe somewhere else,
Changes to parameters are password protected. The password allows access for
30 minutes after entry.