INSTALLATION
3.5
Connections
Fig. 10: Connections
Connections and shielding
Connection cable shield
1
Mark the lines (4 and 5) as intrinsically safe.
2
Connect the shield from (4) to both sides of the EMC cable gland (3).
3
Cable shielding for analog and digital output (5):
– Option 1: Connect the shield as capacitive:
▸
Connect the cable shield to the shielding terminal in the enclosure.
– Option 2: Connect the shield to the enclosure:
▸
Connect the cable shield to the cable gland
TSA151 voltage supply unit
Power supply 24 V PELV DC provided by the customer or optionally available
EMC cable gland
Connection cable: 0.5 / 2.5 / 5 m (shielded, twisted pair)
Analog and digital outputs
Wiring to PELV24 DC, twisted pair
1
4
5
6
2
3
WARNING:
Unallowed voltage supply voids the Ex certification
Only connect the TRANSIC151LP via the TSA151 voltage supply unit.
Only use a 24V DC PELV power supply unit on the TSA151 voltage supply unit.
▸
Observe the “X” marking on the type plate,
see “Product identification”, page 11
WARNING:
Using unallowed cables voids the Ex certification
If the cables are not SICK standard cables, they must meet the cable specifications in
Table
“TSA151 safety-related technical characteristic data (IECEx/ATEX)”, page 76
.
WARNING:
Damage to cabling through heat
▸
Only use cables specified for temperatures >70°C.
NOTE:
Risk of damage to the TSA151 through dust or humidity
▸
Only open the TSA151 in a dry environment free from dust.