7.2
Connection diagrams
Connection diagram for the TMx88
Zero-point setting
Figure 18: Current output connection diagram for the TMx88
Zero-point setting
Figure 19: Voltage output connection diagram for the TMx88
Connection diagram for the TMM55
Figure 20: Current output connection diagram for the TMM55
Figure 21: Voltage output connection diagram for the TMM55
7.3
Length of cable and supply voltage for current output
For sensors with current output, the required supply voltage increases by the voltage
dip on the connected cable. The most significant voltage dip on the cable occurs when
the maximum current of 20 mA is flowing, caused by the resistance of the cable (R
L
).
The partial resistances of the go and return line must also be taken into account.
Figure 22: Length of cable for current output
CONNECTION
7
8019053/1EF6/2021-12-13 | SICK