5.4.2.1
Span points
Menu 2.2.1: Adjustment/manual/Span points
The span point adjustment (recalculate the correction factor) of the selected span point is
manually performed in this menu item.
1
Set the maintenance signal (
2
Activate the “Adjustm. manual” operating state (
).
3
Feed test gas to the gas sampling system (
).
4
Wait for the end of the fill time (if necessary, exit the menu with “
Back
” to review the fill
behavior in the line graph
).
5
Select the desired “Span pt”.
6
Press "
Set
": The measured value is set to the nominal concentration (correction factor
is set).
If the deviation is too large (parameter settings
SOPAS ET), the device switches to
“
Maintenance request
).
7
Press “
Back
” to exit the menu.
8
Go to the “Accept” menu (
) and accept the new values.
9
Reset the maintenance signal (
5.4.2.2
Accept
Menu 2.2.2: Adjustment/Manual/Span points
The determined new span points are set in this menu item.
2.2.1
●
Configured software (
Maintenance
Back
Set
Adjustm. manual
/Adj/Man/Span pt
1 Span pt 1
2 Span pt 2
3 Span pt 3
4 Hg conc
3.0 µg/m3
6.3 µg/m3
9.0 µg/m3
9.3 µg/m3
●
Current measured Hg concentration (measured value)
2.2.2
●
Nominal value
●
Measured value
●
Factor:
Calculated correction factor
Set measured value and correction factor.
Maintenance
Back
Set
Set span pts
/Adj/Man/Accept
1 Span pt 1
2 Span pt 2
3 Span pt 3
4 Factor
3.0 3.0
6.0 6.0
9.0 9.0
1.00000