5.4.1.1
Adjustment/Test/Parameter
Menu 2.1.3: Adjustment/Test/Parameter
5.4.2
Adjustment/manual
Menu 2.2: Adjustment/Manual
Adjustment with gas feeding on the gas sampling system (external test gas) can be started
in this menu.
Test gas connection
2.1.3
●
Enter the concentrations of the respective span point.
Max. 3 concentrations.
Measuring
Back
Set
Parameter
Adj/Test/Par.
1 Span pt 1
2 Span pt 2
3 Span pt 3
4 Span pt 1
5 Span pt 2
3.0 µg/m3
6.3 µg/m3
9.0 µg/m3
Activate
Activate
To change a value:
1
Select the desired span point.
2
“
Enter
”.
3
Change the value. To do so:
Enter the numeric value.
Minus sign: As first digit: Press
several times.
4
To save: “
Save
”.
5
The password is queried (
Measuring
Back
Save
Adj/Test/Par.1
00003.000 µg/m3
●
Activation/deactivation of the span point for the automatic
sequence
2.2
Check and set span points
Measuring
Back
Enter
Manual
/Adj/Man
1 Span points
2 Accept
If a new span point was set in submenu 2.1.1:
Start adjustment here with new span points.