Preparations for Initial Start-up
Su
bj
ec
t t
o c
han
ge
w
ith
ou
t n
ot
ic
e
3.5.3
Test gas feeding on gas sampling system
●
Test gas temperature: 200 °C (test gas feeding via heated sample gas line).
●
Test gas specification
Technical Data.
Connection of test gas line for manual test gas feeding on the gas sampling system:
Fig. 6
Test gas connection
Throughflow during test gas feed: Approx. 500 l/h.
– The device builds up only slight counterpressure.
– Regulate the flow via the pressure.
– Max. pressure: 0.5 bar (50 kPa) overpressure.
After unscrewing the test gas line: Screw the dummy cap back on.
For information on the gas sampling system SFU
“Operating Instructions
SFU”
Connection for manual test gas feeding:
6 mm pipe connection