A D D E N D U M T O O P E R A T I N G I N S T R U C T I O N S | GMS840
Subject to change without notice
3
INSTALLATION
3.6.2
Connecting the power connection
Safety information and standards
▸
Standards to be applied: IEC 60947-1 and IEC 60947-3
▸
Check whether the available power voltage matches the type plate specifications. If this
is not the case: Do not connect the device.
Installing an external power fuse
Install an external circuit breaker in the main power supply.
●
Fuse rating and trigger characteristic:
– Power voltage 115 V AC circuit breaker for 16 Ampere characteristic C
– Power voltage 230 V AC circuit breaker for 16 Ampere characteristic B
Installing an external main power switch
▸
Install a power disconnector switch close to the device (all poles).
▸
Mark the power disconnector switch clearly.
Install the power connection
1
Open the enclosure cover.
2
Insert the power cable through a cable inlet. Use an EMC connection.
3
Lay a shielding on the EMC connection.
4
Connect the power cable to the power connection terminals,
.
5
Close the cable gland on the cable.
6
Connect the external PA connection with the same electrical potential as the internal PE
connection.
WARNING: Health risk
Ensure electrical safety:
▸
Only connect the device to a main power supply which has a functional protective
conductor (PE to PA,
see “Position of connections”, page 29
).
▸
Only start the device when a correct protective conductor is installed.
▸
Do not disconnect protective conductor connections.
Internal power fuses:
●
Primary: Fuse in internal power supply unit (6.3 A)
If this fuse triggers: Replace the power supply unit.
●
Secondary: Fusible cutout on the internal “distribution board” (on CANopen
connections);
if this fuse triggers: Clear the malfunction cause and replace the fusible cutout with
an identical type (F10AH250V).
WARNING: Health risk
▸
Before installing the power cable:
Make sure the external main power supply is
switched off.
Technical requirements to the power cable,
see “Power connection”, page 31
WARNING: Risk of explosion of GMS841 (ATEX/ IECEx) and GMS842 (
C
CSA
US)
▸
Only use “identical” material suitable for PA connections.
▸
Observe the commissioning before switching on the power supply:
“Commissioning of GMS841 (ATEX/ IECEx) and GMS842 (
.