FLOWSIC60
Settings
"Settings" menu with submenus for editing the system
configuration at commissioning and parameterization.
Parameters can only be changed in "Configuration" mode.
Settings
Calibration
Submenu "Calibration" with the basic parameters for the
calibration of the meter according to the field conditions.
Adjust factor
1.0000
Display/Change of the linear correction factor.
Adjust offset
1.0000
Display/Change of the zero-point offset.
Path length
6.2 m
Display/Change of ultrasonic path length. Adjust value to
exact distance between both sensors.
Path angle
60 °
Display/Change of angle between the ultrasonic path and the
main air flow direction. Adjust to the correct value.
Settings
Analog output
Submenu "Analog output" with the parameters of the analog
output
AO at 20mA
10 m/s
Air velocity at maximum current.
AO at 4mA
-10 m/s
Air velocity at minimum current
AO at error
22 mA
Current output in "Error" state
AO damping
0 s
Averaging period for the current output in order to smooth
peaks
Settings
Serial Interface
Submenu "Serial Interface" with the parameters for the serial
interface
Serial Interface
Reset RS485/1?
Resetting the parameter serial interface 1 to the default
values. Defaults are: Baudrate 9600, protocol ASCII
(MODBUS ASCII), 8, n, 1. Master burst mode off.
Serial Interface
Reset Service?
Resetting the parameter of the serial service interface to the
default values (see above)
Settings
LCD Display
Submenu "LCD Display" with language parameter for the LCD
display
Language
English
Set parameter "Language" to
- English
- German
- French
- Russian